NanoDrop Spectrophotometry Protocol

ENTM201L - General Entomology Laboratory | UC Riverside |

NanoDrop DNA Purity Assessment Protocol

Protocol Overview

This protocol uses UV-Vis spectrophotometry to measure DNA purity via A260/A280 and A260/A230 ratios. These ratios detect protein and salt contamination that could inhibit PCR. While NanoDrop provides concentration estimates, trust Qubit for accurate dsDNA quantification and use NanoDrop for purity troubleshooting.


Materials and Equipment

Instrumentation

Reagents

Safety Equipment

Important

NanoDrop uses a Xenon flash lamp that emits UV light. Never look directly at the measurement pedestal when the arm is lowered during measurement. UV exposure can damage eyes.


Part 1: Instrument Initialization

Step 1: Power On and Software Launch

1 Turn on NanoDrop spectrophotometer (power button on rear or side)

2 Launch NanoDrop software on computer

3 Wait for instrument initialization (approximately 30 seconds)

4 Select "Nucleic Acid" module in software

Pro Tip

If the instrument was used previously, pedestals may have dried sample residue. Always start with cleaning step even if pedestals look clean.

Step 2: Clean Measurement Pedestals

5 Pipette 2 µL deionized water onto lower pedestal

6 Lower sampling arm to spread liquid between pedestals

7 Raise arm, wipe both upper and lower pedestals with clean Kimwipe

8 Repeat cleaning cycle 2-3 times until no residue visible

Critical

Residue from previous samples will contaminate your readings. Clean thoroughly between sample sets. Dried guanidine salts (from extraction buffers) appear as white crystals - these must be removed.


Part 2: Blank Measurement

Why Blank?

The blank establishes baseline absorbance for your elution buffer. If your DNA is in TE buffer, blank with TE. If in water, blank with water. Buffer and sample must match or readings will be inaccurate.

Step 3: Prepare and Measure Blank

1 Ensure pedestals are clean and dry

2 Pipette 1-2 µL elution buffer (TE or water) onto lower pedestal

3 Lower sampling arm gently to create liquid column

4 In software, select "Blank" button

5 Wait 3-5 seconds for measurement

6 Software will display "Blank successful" message

7 Raise arm, wipe pedestals with Kimwipe

Pro Tip

Blank only once at the start of a measurement session. Do not re-blank between samples unless you change elution buffers or have a long gap (>30 min) between measurements.


Part 3: DNA Sample Measurement

Step 4: Measure First DNA Sample

1 Wipe pedestals clean with Kimwipe

2 Vortex DNA sample tube for 5 seconds (critical - DNA settles)

3 Quick spin tube to collect liquid at bottom

4 Pipette 1-2 µL DNA onto lower pedestal

Pipetting Technique

Use a P2 pipette set to 1.5 µL for best accuracy. Avoid introducing bubbles. If bubble forms, wipe pedestal and re-pipette. Bubbles scatter light and give false readings.

5 Lower sampling arm to spread DNA between pedestals

6 In software, enter Sample ID (e.g., "Sample1_MagBeads")

7 Click "Measure" button

8 Wait 3-5 seconds for measurement

Step 5: Record Critical Data

Software will display spectral curve (190-840 nm) and calculated values:

Metric What to Record What It Means
Concentration ng/µL (dsDNA) Estimate of DNA amount (compare to Qubit)
A260 Absorbance value Raw absorbance at 260 nm (nucleic acids)
A260/A280 Ratio (unitless) Protein purity indicator (target: 1.8-2.0)
A260/A230 Ratio (unitless) Salt/organic purity indicator (target: 2.0-2.2)

9 Record all four values in lab notebook

10 Optional: Click "Export" to save full spectrum as PDF

Step 6: Sample Recovery (Optional)

11 Raise sampling arm

12 Immediately pipette DNA sample back into original tube

Sample Recovery

NanoDrop is non-destructive. You can recover your 1-2 µL sample if needed. This is useful for low-yield extractions where every microliter counts. Recovery rate: ~80-90%.

Step 7: Clean and Measure Next Sample

13 Wipe both pedestals with Kimwipe

14 Repeat Steps 1-12 for each DNA sample

Avoid Cross-Contamination

Always wipe pedestals between samples. Carryover of just 0.1 µL from a high-concentration sample can contaminate low-concentration readings.


Part 4: Interpreting Results

A260/A280 Ratio Interpretation

Ratio Interpretation Likely Cause PCR Impact
1.8-2.0 Pure DNA Successful extraction Proceed with confidence
1.75-1.8 Slight protein contamination (acceptable for mosquito DNA) Sclerotized cuticle proteins PCR usually succeeds
1.5-1.75 Moderate protein contamination Incomplete Proteinase K digestion Dilute 1:2; add BSA to PCR
<1.5 Heavy protein contamination Lysis failure or insufficient washing Re-extract recommended
>2.0 RNA contamination or very pure DNA No RNase treatment Usually acceptable; RNA doesn't inhibit PCR much

A260/A230 Ratio Interpretation

Ratio Interpretation Likely Cause Action Required
2.0-2.2 Pure DNA, minimal salts Excellent washing None - proceed to PCR
1.8-2.0 Acceptable purity Trace salt carryover Proceed to PCR (monitor for inhibition)
1.5-1.8 Moderate salt contamination Incomplete ethanol removal or guanidine carryover Dilute 1:5 for PCR
<1.5 Heavy contamination Binding buffer carryover; ethanol not evaporated Re-extract with extra washes
>2.2 Very pure or degraded DNA Excellent technique or DNA fragmentation Check integrity on gel

Part 5: Comparing NanoDrop to Qubit

Expected Discrepancies

Scenario 1: NanoDrop Concentration >> Qubit

Example: NanoDrop = 40 ng/µL, Qubit = 15 ng/µL

Interpretation:

  • RNA contamination (absorbs at 260 nm, doesn't bind Qubit dye)
  • Melanin from mosquito eyes (absorbs UV, doesn't fluoresce)
  • Degraded DNA fragments (absorb at 260 nm, bind Qubit poorly)

Action: Trust Qubit for PCR calculations. NanoDrop is overestimating due to contaminants.

Scenario 2: NanoDrop ≈ Qubit, Good Ratios

Example: NanoDrop = 22 ng/µL, Qubit = 20 ng/µL, A260/A280 = 1.88, A260/A230 = 2.05

Interpretation: Pure dsDNA with minimal contamination

Action: Proceed to PCR with confidence. Either concentration value is acceptable.

Scenario 3: Low Ratios Despite Adequate Concentration

Example: NanoDrop = 18 ng/µL, A260/A280 = 1.65, A260/A230 = 1.40

Interpretation: Heavy protein AND salt contamination

Action:

  • Dilute 1:5 for PCR to reduce inhibitor concentration
  • Add BSA to PCR at 0.4 µg/µL final
  • If PCR still fails, re-extract with longer Proteinase K incubation and extra washes

Part 6: Troubleshooting Common Issues

Problem: Erratic or Unrepeatable Readings

Causes and Solutions:

Problem: Negative A260/A230 Ratio

Cause: Absorbance at 230 nm is higher than at 260 nm (severe contamination)

Interpretation: Massive guanidine salt carryover or phenol contamination

Action: Do not use for PCR. Re-extract DNA from tissue.

Problem: A260/A280 >2.5

Cause: RNA contamination or buffer mismatch

Check: Did you blank with the correct elution buffer?

Action:

Problem: Concentration Shows "Negative" or Very Low Despite Visible Gel Band

Cause: Blank was contaminated or sample is very dilute

Action:


Part 7: Post-Measurement Shutdown

Step 8: Clean and Secure Instrument

1 Pipette 2 µL deionized water onto lower pedestal

2 Lower arm, raise arm, wipe both pedestals thoroughly

3 Repeat cleaning 2-3 times

4 Lower sampling arm onto clean pedestal (protects optics)

5 Close NanoDrop software

6 Power off instrument (optional - can be left on)

Maintenance

Always leave the sampling arm lowered when not in use. This protects the optical surfaces from dust and reduces photobleaching of the measurement spot.


Data Recording Template

Sample ID NanoDrop (ng/µL) A260 A260/A280 A260/A230 Quality Assessment
Blank (Buffer) 0.0 0.000 Baseline
Sample 1 _______ _______ _______ _______ _______
Sample 2 _______ _______ _______ _______ _______
Sample 3 _______ _______ _______ _______ _______

Key Takeaways

Critical Concepts

  • NanoDrop measures all UV-absorbing molecules: DNA, RNA, proteins, salts, melanin
  • Qubit measures only dsDNA: Use Qubit concentration for PCR setup
  • Purity ratios are diagnostic: They tell you what contamination is present
  • A260/A280 detects proteins: Low ratio = incomplete Proteinase K digestion
  • A260/A230 detects salts: Low ratio = guanidine or ethanol carryover
  • Mosquito DNA is "difficult": A260/A280 of 1.75-1.85 is normal and acceptable

Best Practice

Use NanoDrop and Qubit together. NanoDrop for purity troubleshooting, Qubit for accurate quantification. Together, they give you complete quality assessment for informed PCR decisions.